Product Overview
The R-series demineralizers produce ultra-pure water that meets the requirements of PN-EN ISO 3696:1999 for first-degree water and EP. They have extensive automation enabling monitoring of work at selected stages of transfer, data archiving, individual settings of alarm thresholds, control of the function of flat membrane connections, and an indication of the content level.
Cat. no: 10DR-TOC-UV
– Fed by tap water.
– Water purification process levels:
• sediment pre-filter 5µm,
• integrated module (sediment-carbon-softening),
• demineralization on an spectrally clean ion-exchange TOC bed,
• reverse osmosis,
• UV lamp 185/254 nm,
• microfiltration capsule 0,45/0,2μm.
– Device efficiency min. 10-12 dm3/h.
– Demineralized water conductivity  0,55 μS/cm.
– Water collection point – I purity class (PN-EN ISO 3696:1999, ASTM, EP).
– Obtained water may be used for instrumental analyses AAS, ICP/MS, IC, HPLC, GC, bacteria cultures, biochemical analyses.
– System equipped with a 10 dm3 pressure storage tank.
– Automatic ultrapure water recirculation between water collections.
– Fed by cold water: 5-40ºC.
– Power supply: 230V/50Hz.
Monitoring functions:
- The device is equipped with a microprocessor 24V automatics, that includes:
• Color display screen with a Touch Panel function,
• Conductometer measuring water pressure: feed, processed by reverse osmosis, and ultrapure (in µS/cm or MOhm),
• Reading values compensated and uncompensated thermally,
• Clock displaying date and time,
• Actual system status information,
• Alarm informing about necessity to replace initial cleaning modules,
• Alarm informing about necessity to replace a UV lamp radiator,
• Alarm informing about necessity to replace RO module,
• Alarm informing about necessity to replace an ion-exchange module,
• Alarm informing about necessity to replace a microfiltration capsule,
• Membrane module retention level information,
• Service dates display,
• Tank fill level,
• Menu in English, Russian, Germany, Spanish or Croatian on display,
• RS 232 connector for service frequency and alarm levels adjustment,
• USB connector for service frequency and alarm levels adjustment,
- Built-in feed water manometer.
Protection functions:
- Automatic pump shutdown when:
• the feed water pressure is too low (lack of feed water) – low pressure sensor,
• the tank is full – high pressure sensor.
- Thermal osmotic module performance protection, automatic system shutdown then the feed water temperature is below 40Cor above 400C.
- Optional system shutdown if any alarm occurs.
- Optional system automatic start.
- System messages/monitoring alarms display.
Required connections in the installation place:
- cold tap water connection ½” or ¾”,
- drainage,
- 230 V power socjet.
Purified water parameters:
- Purified water fits the PN-EN ISO 3696: 1999, the ASTM, and the CLSI standards for the first purity class.
- Obtained water is microbiologically and physicochemically compliant with the FP standard for the purified production water.
- Obtained water may be used for: instrumental analyses AAS, ICP/MS, IC, HPLC, GC, bacteria cultures, biochemical analyses.
Package Included:
Sediment prefilter 5µm EO-005-10
Module A2 EO-MA-12
Module H7 TOC* EJ-2000-1
Module H6 TOC* EJ-5000-1
Capsule microfiltration EM-SP-20
UV lamp 185/ 254 nm** EUV-185-254-0
| Product More Specifications | |
| Made In | Poland | 
- Stock: Pre Order
- Model: Hydrolab
- SKU: 10DR-TOC-UV
 
 
 
 
 
 
 
 